Product Name |
Volumetric Infusion Pump |
Model |
MEDIFUSION DI-2000 |
MEDIFUSION DI-2200 |
Pumping Mechanism |
Peristaltic Transit Finger |
Flow Rates |
0.1 ~ 1200 ml/h |
Micro(on) : 0.1 ~ 99.9 ml/h(in 0.1 ml/h increments) 100 ~ 1200 ml/h(in 1 ml/h increments) Micro(off) : 1 ~ 1200 ml/h(in 1 ml/h increments) |
Accuracy |
±5% with assigned IV set Measured every hour at room temperature(25°C) using water or saline solution(above 1.0 ml/h of flow rate). |
Infusion Range |
Total Vol. |
0.0 ~ 9999 ml or infinity |
Micro(on) : 0.1 ~ 99.9 ml/h(in 0.1 ml/h increments) 100 ~ 9999 ml/h(in 1 ml/h increments) Micro(off) : 1 ~ 9999 ml/h(in 1 ml/h increments) |
Infused Vol. |
0.0 ~ 9999 ml |
0.1 ~ 99.9 ml(in 0.1 ml increments) 100 ~ 9999 ml(in 1 ml increments) |
Bolus Rate |
700 ml/h(default) - 1 ~ 1200 ml/h |
Bolus Volume |
3 ml(default) - 1 ~ 9999 ml |
Purge Rate |
700 ml/h(default) - 1 ~ 1200 ml/h |
Purge Volume |
3 ml(default) - 1 ~ 9999 ml |
KVO Rates |
1 ml/h(default) 1~9 ml/h |
At Flow Rate 1.0 ~ 1200 ml/h |
0.1 ml/h |
At Flow Rate 0.1 ~ 0.9 ml/h |
Display |
7 segments(4 digits x 3 lines) |
Power |
* AC 100~240 VAC, 50~60 Hz(Fuse : 250V, T3.15A) * DC 12(500mA) * Ni-MH Rechargeable Battery - 2100 mA/h, AA, 1.2V, 10EA - operating time : approx. 4+ hrs at a flow rate of 25ml/h - recharging time : approx. 10 hours |
Power Requirements |
35 VA |
Protection Type |
* Protection from electric shock : Class I * Protection capacity for electric shock : Type CF * Protection from liquid infiltration : IPX 1(drip-proof)
|
Size(D x W x H) |
120 x 130 x 206mm / 120 x 215 x 206mm |
Weight |
approx. 1.7kg / 3.0kg |
Operating Environment |
Temperature : 5~40°C, Humidity : 20~90% Atmospheric Pressure : 70~106 kPa(525.04~795.06mm Hg) |
Storage Environment |
Temperature : -10~60°C, Humidity : 10~95% Atmospheric Pressure : 65~120 kPa(487.54~900.07mm Hg) |
Accessories |
AC power cable(4m, 0.75mm2 x 3C), Infusion pole clamp(A type), Operation manual-one piece each, Drop Sensor(A type) |
Communication |
USB, RS485(optional) |